Posey Company • 5635 Peck Road, Arcadia, CA 91006-0020 USA • www.posey.com
Phone: 1.800.447.6739 or 1.626.443.3143 • Fax: 1.800.767.3933 or 1.626.443.5014
MDSS GmbH
Schiffgraben 41
D-30175 Hannover, Germany
EC REP
ApplicAtion
instructions
1. Wrap the limb holder cuff around the
patient’s wrist so the metal ring is
on the ulnar side of the wrist
(opposite the thumb) (fig. 1).
2. Secure the hook and loop fastener.
Slide ONE finger (flat) between the
cuff and the inside of the patient’s
wrist to ensure proper fit (fig. 2). e
strap must be snug, but not
compromise circulation.
3. Bring both connecting straps around
the limb, one clockwise and one
counterclockwise, and feed them
through the metal ring.
4. Place the straps together and attach to a single point along
a movable part of the bed frame, out of the patient’s reach,
with quick-release ties.
5. Adjust the bed strap(s) to allow desired freedom of
movement, without compromising patient or caregiver
safety.
NOTE: To remove cuffs, untie the straps and release hook and
loop fasteners.
Applicable Products:
REF
2520, 2528
DEscription oF proDuct: Limb holder for limiting limb movement. For bed use only.
rx onlY
liMB HolDErs
© 2009 Posey Company. All rights reserved.
I9215 062209
2520
posEY liMB HolDErs
REF
2520 Double Strap Foam Limb
Holders - 3/8” (1 cm) foam
REF
2528 Double Strap Foam Limb
Holders - 1/4” (5/8 cm) foam
ApplicAtion instructions:
Fig. 2
ADDitionAl sAFEtY AnD lAunDEring
instructions on otHEr siDE
inDicAtions For usE:
• Patientsassessedtobeatriskofdisruptinglife-savingtreatments(e.g.,chronictubepulling).
• Patientsassessedtobeatriskoflinepulling,whichmaypreventmonitoringofvitalsigns.
• Patientswhosepicking,pulling,scratching,orpeelingexacerbatesaskincondition,causesself-
injury, or compromises wound site integrity.
ADVErsE rEActions
• Severeemotional,psychological,orphysicalproblemsmayoccur:iftheapplieddeviceis
uncomfortable; or if it severely limits movement. If symptoms of these problems ever appear
for any reason, get help from a qualified medical authority and find a less restrictive, product or
intervention.
contrAinDicAtions:
• DO NOT use on a patient who is or becomes highly aggressive, combative, agitated, or
suicidal.
MONITOR PER FACILITY POLICY. Check to ensure that:
• Strapscannotslideinanydirectionorloosenifthepatient
pullsonthem,orifthebedisadjusted;
• Cusareproperlysecured.Ifappliedtootightly,
circulationwillberestricted;ifappliedtooloosely,the
patientmaybeabletosliphisorherlimbfromthedevice;
• Cusareattachedinawaythatthepatientisnotableto
usehisorherteethorotherwiseremovethedevice;
• Cusareintact,andnottornordamaged.DONOTallow
patientstoingestproductmaterial.
ADDITIONALORDIFFERENTBODYORLIMBRESTRAINTS
MAY BE NEEDED:
• Ifthepatientpullsviolentlyagainstthebedstraps;
• Toreducetheriskofthepatientgettingaccesstotheline/
wound/tubesite;
• Topreventthepatientfromailingorbuckingupand
downcausingself-injury.
SeePoseyCatalog.
BEDSAFETY
• ALWAYSuseHospitalBedSafety
Workgroup(HBSW)(http://www.fda.
gov/cdrh/beds/modguide.html)
compliantsiderailsintheUPposition
andllALLgapstoreducetheriskof
entrapment.
• Usesiderailcoversandgapprotectorstohelppreventthe
patient’sbodyfromgoingunder,around,throughor
betweenthesiderails.Afailuretodosomayresultin
seriousinjuryordeathifapatientbecomessuspendedor
entrapped.Poseyoersafullrangeofsiderailpadsand/
orgapprotectorstocovergaps.
Before use, check device for damage. Discard if you
have any questions about patient safety.
• DONOTtieaknotatthemetalring.eknotwouldbe
diculttountieinanemergency.
• NEVERseparatethestraps,ortriangulatethisproduct.If
thepatientpulls,thiswouldallowtherestrainttotighten
onthelimbandrestrictcirculation.
ADDitionAl WArnings:
• NEVERuseonapatient:
•
With a dislocation or fracture on the restrained limb; or
•
If an IV or wound site could be compromised by the device.
See the Posey Catalog for other options for such a patient.
Fig. 1
Comments to this Manuals