Application
Instructions
DESCRIPTION OF PRODUCT: Straitjacket with full length sleeves which secure behind the patient.
Rx ONLY
Straitjacket
© 2012 Posey Company. All rights reserved.
I9236 REV B 030712
ADDITIONAL SAFETY AND LAUNDERING
INSTRUCTIONS ON OTHER SIDE
• DO NOT use on patients with: ostomy, colostomy, or G-tubes;
hernias, severe Cardio Obstructive Pulmonary Disease (COPD);
or with post-surgery tubes, incisions or monitoring lines.
These could be disrupted by a restraint.
8118
Posey Straitjacket
REF
8118 Straitjacket, S, M, L, XL
• NEVERallowarestrainedpatienttowalkwithout
assistance.Thepatientisatriskofinjuryfromafallor
harmfromothers.
Application Instructions:
Adverse Reactions:
• Severeemotional,psychological,orphysicalproblemsmay
occur: if the applied device is uncomfortable; or if it severely
limits movement. If the patient is likely to cause injury to
him/herself or others, get help from a qualified medical
authority to find an alternate intervention or method of restraint.
Indications for Use:
• Patientsassessedasbeinginextremedangerofinjuryto
themselves or to others.
Contraindications:
MONITORING
• Poseyrecommendsconstantdirectsupervisionforpatients
deemedtobeatriskofinjurytothemselvesorothers.For
timeswhendirectsupervisionisnotpossible,monitorby
lineofsightorbyavideo/audiodevice.
• NEVERallowapatienttohaveaccesstoanytool,utensil,
orobjectthatmightbeusedtounlockordamagecuffs.
• Beawarethatasuddenmoodswingmaycauseagitatedor
aggressivebehavior.ContactthemedicalteamATONCEif
thisoccurs.Nolevelofmonitoringmayavoidtheriskof
seriousinjurytohighlyagitatedoraggressivepatients.
• NEVER use on a patient:
• With a dislocation or fracture on the restrained limb; or
• If an IV or wound site could be compromised by the device.
See the Posey Catalog for other options for such a patient.
• Beforeeachuse,checkfortears,frayingorexcessivewear;
orcrackedorbrokenbucklesorlocks;and/orthathook
andloopadheressecurely.Discardifdeviceisdamaged.
• Ifyouhaveanyquestionsaboutpatientsafety,askthe
appropriatemedicalauthorityforalternatives.
• Makesurepatientwearsproperundergarmentstoprotect
skin.
• DO NOT use this device on a patient who is or becomes:
suicidal; highly aggressive or combative; self-destructive;
or deemed to be an immediate risk to others, UNLESS the
patient is under constant supervision.
1. With the opening at the back, put the patient’s arms through
the sleeves.
2. Close the back by feeding the ends of the straps through each
of the four corresponding buckles.
3. Cross the patient’s arms in front, and pass each arm through
the vertical loop at chest level.
4. Pass each arm through the vertical loops located on the sides
of the jacket underneath the arm.
5. Secure the two sleeve ends behind the patient’s back using the
attached straps.
6. Pass the pelvic strap between the patient’s legs and secure
with the buckle behind the patient’s back.
7. Check all straps to ensure they are snug enough to prevent
escape, but do not interfere with circulation or breathing.
• Checkthepatient’slimbs,circulationandbreathing
frequently.Ifthisdeviceistootight,circulationor
breathingwillberestricted.Iftooloose,thepatientmay
beabletoescape.